Part
VI
Building your own
Introduction
Building your own photon counting system has several advantages. First
is the satisfaction of creating a sensitive instrument capable of
detecting individual photons. You will also have a first hand knowledge
of the instrument so repairs can be easily made. Most units seldom
need repairs, however. Lastly, by building your own photometer system
you can save considerable money.
Figure 8 shows a block diagram of a basic astronomical photon counting
system.

Figure 8
Astronomical Photon Counting System Block Diagram
Figures 9 and 10 show close-ups of the photometry equipment built
and used at the Hopkins phoenix Observatory.

Figure 9
Telescope and Photometer Head

Figure 10
10 MHz Pulse Counter and -1,000 VDC Power Supply
Photometer
Head
Building your own photometer head can be rewarding. Someone once said
you will build at least three before getting one that you are happy
with. Hopefully the following information will make it easier for
you.
Figure
11 shows a basic photometer head. Many variations of this basic design
can be made.

Figure 11
Basic Photometer Head
The
basic requirement for a photometer head is that the photomultiplier
tube must be in the light path of the object being measured. Because
of the photocathode surface variations, a Fabry Lens is used to spread
the light on the photocathode and minimize the effects. A diaphragm
is used to stop down a star and minimize background light from the
sky and other stars. The diaphragm must have several openings ranging
from a large opening for initially centering the star to the smallest,
usually 60 arc seconds. To center the star in the diaphragm, an illuminated
dual-cross hair post-view eyepiece and first surface flip mirror are
required. A flip prism can be used in place of the flip mirror. A
pre-viewer eyepiece and flip mirror is optional, but makes finding
the star initially easier. A good finder scope on the main telescope
can serve this purpose better than the pre-viewer, however. For faint
objects beyond the capability of the finder scope, the pre-viewer
allows use of the main telescope's aperture for finding the object.
Lastly, for filter photometry a filter must be in the optical path
just before the photocathode. Since various filters are normally used,
a means of easily changing the filters must be made.
How
the items are packaged is limited only by your imagination.Large
telescopes can allow generous use of volume and weight of the head.
Telescopes on the order of the popular 8" SCTs must have a more compact
and light weight design. To create such a light weight and compact
photometer head more clever designs are needed.
The
addition of first surface mirrors can be used to fold the light path
to the photomultiplier tube. Prisms and first surface mirrors can
be used to change the light path for the post viewer. Only first surface
mirrors, never prisms, should be used in the light path between the
star and detector. Aluminum is the choice of materials for the housing
as it can provide a strong light weight and electromagnetic shielded
enclosure. All aluminum should be black anodized and/or painted flat
black inside and out.
In
the early 1980s I designed and tested several photometer heads. I
settled on a very compact and light weight design. The design uses
a 4.7" X 3.7" X 2.1" die case aluminum box (BUD #CU-234). While the
box is close to ideal, the sloping sides make working with it difficult.
I had a machinist friend do the milling and drilling work for me.
For those skilled with access to a milling machine, the whole head
could be milled out of a solid block of aluminum.
Figure
12 shows a diagram of the compact photometer head designed and used
at the Hopkins Phoenix Observatory. The photometer head is about the
size and weight of a typically 35 mm camera.

Figure 12
Photometer Head Top View Drawing
Figure
13 shows a photograph of the inside of the photometer head. The case
is aluminum that has been black anodized. There is a lip around the
cover which provides an excellent light seal. A small screw at the
middle left side of the case is used to align the fixed mirror for
the photomultiplier tube (the tube assembly is removed during this
alignment). When done, the screw is fixed in place. The filter cage
and diaphragm use screw type detents to key positions. The rods for
the flip mirror and fixed mirror are 1/8" brass brazing rods painted
black. Note the filter cage is in the dark slide position.

Figure 13
Photometer Head Inside
The
Fabry lens is mounted on a bracket on the cover. See Figure 14. When
the cover is in place the lens is in position between the filter cage
and fixed mirror.

Figure 14
Fabry Lens Mounting
Figure
15 shows a close-up of the UBV photomultiplier tube photometer head
used at the Hopkins Phoenix Observatory. It's shown mounted on a C-8
telescope. The white box on the back of the photometer head holds
2-AA batteries for the post-viewer illuminated reticle eyepiece. The
knob on the white box varies the brightness of the reticle illumination.
The top knob on the black case is for the flip mirror. The bottom
knob on the black case is the 4-position filter selector (UBV and
Dark). The two RG-58U high voltage and signal cables can be seen out
the bottom (opposite side from the knobs).

Figure 15
UBV Photometer Head
Photomultiplier
Tube
The photomultiplier tube detector is the key to the system. The tube
produces small current pulses when illuminated by photons. The most
popular low cost side-window photomultiplier tube is the 1P21. There
have been several manufactures of these, including RCA and Hamamatsu.
The
931A (usually the lowest cost tube) is identical to the 1P21 except
the 1P21 is a selected 931A. EMI Gencom, Inc. made an equivalent side-window
photomultiplier tube, the EMI 9781. The EMI Gencom, Starlight-1 Photometer
uses an EMI 9924A tube. The EMI 9924A is an on-end window type tube
which can complicate a compact design.
Note:
Many 931A tubes are just as good and some are better than some 1P21s.
To play safe, however, it is best to spend a few extra dollars and
specify a 1P21. If you can afford it, buy several tubes and experiment
to select the best tube. Remember, sometimes tubes may exhibit high
dark current when first turned on, but after sitting in the dark for
several hours with power applied, the dark current may go down to
an acceptable level and may even remain there when first turned on
next time, assuming no exposure to bright light in between. Tubes
that at first may seem less than acceptable may prove fine. Once you
have found a tube, it can last you a life-time.
Other
side-window photomultiplier tubes with different spectral responses
can be used or special projects. A 1P28, for example, extends the
sensitivity further into the ultraviolet region, but at the cost of
much higher dark current. For UBV work, where matching to the standard
UBV system is desired, a 1P21/931A is recommended
Figure
16 shows a typical spectral response curve of the 1P21 photomultiplier
tube.

Figure 16
1P21 Spectral Response
The
1P21 and equivalent photomultiplier tubes require a power supply capable
of producing at least -1000 VDC @ 1 - 2 mA. It has been found, for
analog operation (DC amplifier), voltages from -600 VDC up to -1000
VDC can be used. Once a voltage is determined, it should always be
the same for future observations. If changed, the system zero points
will change. Never change the high voltage in the middle of an observation
set except to correct a drifting voltage. The different sensitivities
will make the data inaccurate. The higher the voltage, the higher
the gain of the tube (also the higher the dark counts). For photon
counting a constant setting of -950 VDC seems best. At 50 degrees
F and -950 VDC typical dark current/count is around 50 counts per
second.
The
typical gain of a 1P21 tube is over 10^6. For DC systems the output
current is in nanoamperes. For photon counting the pulses are in the
low microvolts with low nanosecond pulse widths and sub-nanosecond
rise and fall times. The voltage and times are with the signal terminated
in 50 ohms. High impedance terminations greater than 50 ohms will
lengthen the pulse width. That is undesirable as it will increase
the dead time.
Figure
17 shows a picture of a 1P21/931A photomultiplier tube along with
a drawing of dimensions.
Note:
The tube's photocathode is aligned with the tube socket key (between
pins 1 and 11).

Figure 17
1P21 Photomultiplier Tube
For
new tubes see Hamamatsu at
http://usa.hamamatsu.com/
For surplus tubes see:
http://www.elexs.com/7Photo.htm
http://www.tubesandmore.com/
http://www.vacuumtubes.net/
http://www.vacuumtubes.com/
http://www.vacuumtubesinc.com/
http://www.tubeworld.com/tubes.html
http://www.surplusshed.com/new.html
Voltage
Divider
One of the secrets of a good photometer head, with a photomultiplier
tube, is the voltage divider design. Figure 18 shows a voltage divider
configuration that performs well with a 1P21. It can be used with
either photon counting or analog systems.

Figure 18
Voltage Divider Schematic
If only analog systems are used, the pulse forming capacitors (lower
four 100 WVDC capacitors) can be eliminated. They are very important
for photon counting, however. The 0.001 uF 1000 WVDC capacitor at
the high voltage input is important and must be there. Note that the
voltage rating of the pulse forming capacitors is only 100 WVDC. This
is because the voltage between stages is less than 100 volts. The
smaller capacitors with lower working voltages can therefore be used.
The
voltage divider provides a graduated voltage to all the electrodes
in the photomultiplier tube. When constructing the voltage divider
it is important to make all leads as short as possible. Low nanoseconds
relates to frequencies greater than 100 MHz, so high frequency RF
techniques should be employed. A special low leakage 11 pin photomultiplier
tube socket can be obtained from Hamamatsu Corp.
Socket
A high
quality low leakage tube socket should be used. The 1P21 uses an 11
pin socket. Figure 19 shows a diagram of the tube socket with dimensions.

Figure 19
1P21 Tube Socket
Note:
The tube's photocathode is aligned with the tube socket key (between
pins 1 and 11).
These
sockets an be obtained from Hamamatsu Corp.
They are: Side-Window PMT Tube Socket Part number E9678-11A
Hamamatsu Corp. 420 South Avenue Middlesex, NJ 08846 (202) 469-6640
Magnetic
Shield
The gain of the photomultiplier tube is very sensitive to magnetic
fields. It is even affected by the Earth's magnetic field. Just re-orienting
the photometer head in the Earth's magnetic field (while the photometer
is moved from one position to another) produces a significantly different
gain. Stray fields from power lines, motors, relays, etc. can all
have sever effects on the gain. A simple solution is to use a magnetic
shield. Magnetic shields have been specifically designed for photomultiplier
tubes and can be purchased from most photomultiplier tube manufacturers.
Figure
20 shows the mechanical details of a magnetic shield for a 1P21. These
shields are important and should not be left out. Make sure the opening
in the shield is aligned with the photomultiplier tube's photocathode
(same side as the socket key for the tube). The small holes at the
bottom of the shield is used to lock the shield to the photomultiplier
tube so it doesn't turn. I tapped them and used a small 6-32 nylon
screw for this.

Figure
20
1P21 Photomultiplier Tube Magnetic Shield
Magnetic shields can be obtained from Hamamatsu Corp.
They are: Side-Window PMT Magnetic Shield Part number E989
Hamamatsu Corp. 420 South Avenue Middlesex, NJ 08846 (202) 469-6640
Mirrors
The fixed and flip mirrors should be first surface mirrors. These
can be attached to small brass plates soldered to rods using a small
amount of RTV adhesive.
Pre-Viewer (Optional)
The pre-viewer uses a normal eyepiece typically 25 mm or longer focal
length. It's best to focus the telescope with the post-viewer first.
Then slide the pre-view eyepiece in or out to focus it. Pre-viewers
are good for faint star work where the stars cannot be seen in the
finder. They add complexity and weight to the photometer head, however.
Post-Viewer
To adjust the position of the star in the diaphragm, a post-viewer
is needed. Usually a 12 mm illuminated dual-cross hair reticule eyepiece
is used. The telescope's optics should provide a focus of the star
at the diaphragm. To provide a focus on the diaphragm hole a Relay
Lens is needed. To set the focus of the post-viewer eyepiece, with
power off and the back cover of the head off (make sure the filter
cage is in the dark slide position), illuminate the diaphragm and
focus on the edge of one of the holes. After finding a star and centering
it, focus the telescope so the star is in focus in the post-viewer
eyepiece. This will put the star in focus at the diaphragm.
Relay Lens
By making a relay lens holder that fits inside the bottom of the eyepiece,
the exact focus of the Relay Lens is not critical. The eyepiece can
be moved in and out of the photometer head as well as the relay lens
holder moved in and out of the eyepiece. Once adjusted, the relay
lens can be held in place with a bit of silicon adhesive. The eyepice
can be locked down with a set screw. Figure 21 shows a layout for
a relay lens holder for a 1 1/4" eyepiece barrel.

Figure 21
Relay Lens Holder
Edmund Scientific relay lens part numbers 95034 and 94749.
Edmund
Scientific Co.
101 East Gloucester Pike Barrington, NJ 08007-1380
(609) 547-3488/573-6250
Diaphragm
The purpose of the diaphragm is to mask out all but a small part of
the sky around the star. This has two functions. First, it cuts down
the sky or background readings considerably. Second, it allows a single
star to be selected for measurement out of a field of close stars.
Typical diaphragm sizes are: maximum opening, 120, 90 , and 60 arc
seconds. Some people, with very accurate mounts and well sheltered
from the wind, use 30 and even 15 arc second diaphragms. Usually the
larger the diaphragm you can use the better. This reduces the time
spent with lost data if the star moves out of the diaphragm. Figure
22 shows a diaphragm plate used with the Hopkins Phoenix Observatory
photometer head. The plate has been machined from a 1/16" aluminum
plate and black anodized.

Figure 22
Diaphragm Plate
Making
small diaphragms holes can be a problem. A number 80 drill (0.0135")
will make about a 35 arc second hole when used with an 8-inch f/10
telescope. Longer focal lengths for the same size hole will produce
a smaller angular opening. To calculate the angular size, use the
following:
They
are available from Schott Glass Technologies, Inc.
SCHOTT GLASS TECHNOLOGIES, INC.
400 York Avenue Duryae, PA 18642 (717) 457-7485 http://www.schott.com/optics_devices/english/products/filter/glass_filter.html
and
http://www.andcorp.com/Web_store/Filter_Glass/filter_glass.html
and http://www.omegafilters.com/index.php?page=omegatext/prod_astro_index&ps_session=10aadfa8008db1459b8450bb166380bb
Note:
The UG-5 is more sensitive than the UG-2 filter, has higher ultraviolet
transmission, but is more expensive and has more red leak than the
UG-2. The UG-5 was originally recommended by the AAVSO. The UG-2 is
used for UBV photometry by Kitt Peak National Observatories. A GG-495
filter can be used with the U filter to measure the red leak. This
can be important for "red" stars which have little ultraviolet light.
Since the spectral response of the 1P21 tube drops to near zero in
the region the red region passed by the UG-2 , normally there is little
concern about red leak.
If
there is sufficient interest, filters can be purchased, cut and distributed
at a significant savings. These filters are supplied from Schott in
2.0 inch square glass with an average price of around $50 each (1980
price). Since only 0.5" square filters are needed, a set of 8 filters
can be made from each original 2" square filter. That assumes no breakage.
For more information, contact Jeff Hopkins at
phxjeff@hposoft.com.
Figure 24 shows typical transmission curves for these filters.

Figure 24
UBV Filters Transmission Spectral Curves
Filters
for use in the Red and Infrared bands are as follows:
R - Corning 3480 + 4600 (2 mm thick)
I
- Corning 2600 + 3850 (2 mm thick)
Note:
These filters will not work with the 1P21 as the spectral response
of the tube drops close to zero in longer wavelength bands. A special
photomultiplier that is sensitive in the infrared region is needed
with these filters.
Techniques
For Mounting Filters
Mounting the filters can be done in several ways. The basic criteria
is that the filters must be in the light path after the point of light
deflection to the post-viewer eyepiece. Since visible light will be
nearly completely attenuated with the U filter, looking through it
will not work. In addition, by placing the filters close to the photomultiplier,
the size of the filters can be reduced. When changing the filters
the positioning should be smooth and accurate as well as light tight.
Any
technique that allows smooth changing and light tight operation is
acceptable. There should be one position that is blocked, no filter
and no opening, to provide a "dark" position for when the photometer
head is not in use. This will protect the photomultiplier tube from
stray light. Whatever technique is used the filter mount should be
black anodized and/or painted flat black on all sides.
Slides
The filters can be mounted on a slide. It is very important that the
slides be light tight. The Optec photometers use this method. The
original AAVSO Manual for Astronomical Photoelectric Photometry
describes a photometer head using a slide for the filters.
Wheels
Filter wheels are very popular with CCD cameras,. However, a filter
wheel used for the photomultiplier tube system must be positioned
after the post-viewer eyepiece. This can complicate the implementation.
The operation must also be light tight. Customized units can be made,
however.
Cages
A filter cage that just fits around the magnetic shield can provide
a very compact and light tight arrangement for the filters. The finished
unit should be black anodized and can be machined out of solid aluminum.
Small, 1/2" filters can be attached with a dab of silicon adhesive.
Detents at the top can provide precise stops for the filters. See
Figure 25 for one of the filter cages used at the Hopkins Phoenix
Observatory.

Figure 25
Filter Cage
Cooling
Cooling a photomultiplier tube will increase its sensitivity and reduce
noise. People have devised many different means to cool them, e.g.,
dry ice, cold water and liquid nitrogen. Cooling complicates the design
considerably and the increased bulk makes most cooled systems unsuitable
for small telescopes. For bright star work, cooling is not needed.
Even fainter star work does not require cooling if the ambient temperature
is not above 50 degrees F and/or a large telescope is being used.
High Voltage
Power Supply
Introduction
A photoelectric photometry system, using a photomultiplier tube, requires
a high voltage power supply. One can be constructed easily and inexpensively.
Typically the tube uses -500 VDC to -1000 VDC at around 1 - 2 mA.
The current requirement is a function of the tube's voltage divider.
While it was not too long ago that a high voltage power supply would
have required the use of vacuum tubes, the following circuit provides
an easy to build and inexpensive solid state high voltage power supply.
The basic requirements for the high voltage power supply are an adjustable
negative high voltage of - 500 VDC to -1000 VDC with at least 2 mA
of current available and regulation to at least 0.1%. The following
is a description of a circuit that meets these requirements.
Note:
Because
temperature can affect the power supplies regulation, it is best to
turn the power supply on and let it stabilize for at least 30 minutes
in addition to monitoring the voltage and adjusting any variation
during the photometry measurements.
There
are three parts to the high voltage power supply, the low voltage
section, the oscillator section, and the high-voltage section. Figure
26 shows a schematic of the high voltage power supply.

Figure 26
High Voltage Power Supply Schematic
Low Voltage Section
The low voltage section consists of a bridge rectifier (BR1), a 1000
uF 35 WVDC filter capacitor (C1), an adjustable three terminal voltage
regulator LM317 (VR1), a 120 ohm resistor (R1), 1k ohm potentiometer
(R2), and capacitor C2. Either 12 VAC or 12 VDC can be used. If AC
is used, BR1 provides rectification. If DC is used BR1 corrects for
the proper polarity (either input terminal can be positive). C1 provides
filtering of the input voltage. VR1 allows adjustment of the output
voltage from about 3 VDC to 10 VDC. R1, in conjunction with R2, provides
a feedback voltage for the regulator. By adjusting R2 the output voltage
can be varied. The regulator provides better than 0.1% regulation.
This is line regulation. There is no load regulation and it is not
necessary because the photomultiplier tube's voltage divider provides
a constant load (about 1 milliampere) several orders of magnitude
greater than the variations due to the photomultiplier tube current
(typically less than a microampere). C2 provides transient protection
for the regulator. The voltage adjust potentiometer R2 can be removed
from the board and a panel mounted multi-turn potentiometer used.
Just connect the three wires from where R2 was to the panel mounted
potentiometer.
Oscillator
Section
The oscillator section consists of Timer (8 pin Dip) integrated circuit
U1 (CA555), drive transistor Q1 (2N2219), timing components resistor
R1 (1k ohm) and potentiometer R2 (100k ohm multi-turn) and capacitor
C1 (0.001 uF), bypass capacitors C2 and C3 (0.1 uF), and current limiting
resistor R3 (330 ohm). U1 provides a 30 kHz pulse whose amplitude
is proportional to the supply voltage. The potentiometer R2 is adjusted
for an output frequency from U1 of around 30 kHz. The exact frequency
is dependent on the transformer. The frequency is adjusted for a peak
in the output high voltage with a corresponding dip in the input supply
current. This is a point of resonance for the transformer and provides
the best operation point. Once set, it should not need further adjustments.
The high voltage output is then proportional to the amplitude of the
primary pulse. Adjusting the supply voltage changes the amplitude
of the primary pulse and thus the output high voltage. U1 is not capable
of driving the transformer directly so drive transistor Q1 is used.
The resistor R3 provides a current limited pulse to the base of Q1.
The supply voltage is connected to one side of the primary winding
of transformer T1. When Q1 is biased on, due to a pulse on its base,
it goes into saturation and effectively grounds the other primary
lead of T1. This causes a high current pulse to flow in the transformer's
primary circuit. The primary of T1 should be 4 turns of number 26
enamel coated wire. The primary should be wound first with an insulating
layer of mylar on top.
High
Voltage Section
The high voltage section consists of the secondary winding of transformer
T1, four high voltage diodes forming a bridge rectifier CR1 - CR4
(1N4007), three filter capacitors C1, C2, and C3 (0.001 uF, 1000 VDC),
voltage divider network consisting of resistor R1 (10 megohm) and
potentiometer R2 (100k ohm multi-turn), and an optional digital panel
meter (X-34). The secondary of T1 consists of three layers of number
36 enamel coated wire (with mylar tape insulation between each layer
and each layer having 150 turns) totaling about 450 turns. The exact
number is not important. It is better to have a few extra than not
enough, however. Diodes CR1 - CR4 make a bridge rectifier and capacitors
C1 - C3 provide the filtering.
Low
Voltage Power Supply
Introduction
The pulse conditioning circuit requires low voltage power of +5 VDC
@ 50 mA, -5 VDC @ 50 mA, and +12 VDC @ 20 mA. The 10 MHz Pulse Counter
requires +5 VDC @ 350 mA. The high voltage power supply requires 12
VDC/AC at about 500 mA. To supply these power requirements the following
Low Voltage Power Supply can be used.
Circuit
Figure 27 shows a schematic of a low voltage power supply that can
be used to power the Pulse Conditioner, High Voltage Power Supply,
and 10 MHz Pulse Counter. Three terminal regulators are used to provide
regulated +5 VDC, -5 VDC, and + 12 VDC at up to 1 Amp each. A -12
VDC circuit is included but not needed for the above requirements.

Figure 27
Low Voltage Power Supply Schematic
It
is very important to use heat sinks with these regulators. This may
be done by adding special heat sinks to each regulator or using the
enclosure they are mounted in as the heat sink. If the enclosure is
used, be sure to use heat sink grease and insulating parts (e.g.,
a mica insulator and insulating feed-throughs and washers) to electrically
isolate the regulator from the enclosure. If individual heat sinks
are used the insulating parts are not required but the grease is.
Parts can be purchased from most electronics distributors and as shown
on the schematic. Typically the regulators are under $2 each and the
transformer around $10.
Data
Output
Analog
While the output of a photomultiplier tube is a series of current
pulses, they can be integrated or averaged with a small capacitor
to produce a current that is proportional to the input light. A current
amplifier with several switchable stages of amplification is used.
Display
Meter
A simple analog microampere meter can be used to read the current
from a DC Amplifier. Getting precise reading is difficult, however.
A digital meter would definely be a better choice.
Chart
Recorder
A step
up from the meter is a chart recorder. This was popular for a while.
It not only gave a permanent record of the data, it also allowed a
bit more accuracy than the meter. On the downside, you could easily
go through a lot of chart paper and keeping the data would take non-insignificant
amount of space.
Voltage-to-Frequency
Converter/Counter
A trick to make reading the data more precise, is to convert the current
to a voltage and use a voltage-to-frequency converter, e.g., the Analog
Devices AD537. This allows a greater range and more precise readings.
A frequency, as with photon counting, can be used to display the frequency
or counts. This is not the same as photon counting , however, and
does not provide the dynamic range of photon counting.
Pulse
Counting (Photon Counting)
As noted above the output of the photomultiplier tube is a series
of current pulses. Instead of averaging those pulses and reading the
current, the individual pulses can be amplified and counted. This
allows more dynamic range, sensitivity and perhaps easiest to use
astronomical photoelectric photometry system. The photomultiplier
tube photon-counting system provides at least 2 magnitudes more sensitivity
than a solid state system plus a dynamic range of over 10^7. In addition,
the linearity is excellent. With the large dynamic range, no scale
or gain switching is needed. This allows very accurate work to be
done using a comparison and program star that differ greatly in brightness.
Data reduction is also easier as there are no gain factors to worry
about and reading are precise to 6 or 8 decimal positions. Correction
for the system dead time must be made, however.
The
basic requirements for a photon-counting system are a detector (e.g.,
a photomultiplier tube) that produces current pulses in proportion
to incident photons, some sort of pulse conditioner, low voltage power
supply, and a pulse or frequency counter. The output pulses from a
photomultiplier tube are negative going with an amplitude typically
in the microvolt region. The pulse width is typically less than 10
nanoseconds with rise and fall times in the sub-nanosecond range.
The voltage and times are with the signal terminated in 50 ohms. Higher
resistance termination will result in wider pulses and correspondingly
larger dead time. The pulse conditioner must take these fast low-level
pulses and convert them into something that can trigger a frequency
counter or a computer counter circuit.
A
counter capable of at least 10 MHz should be used. Bright star work
can result in counts into the millions for a ten-second integration.
Counts above that should be avoided. With larger telescopes and brighter
stars, the photomultiplier tube may become saturated. When saturated
the tube loses linearity and should be avoided. It may be necessary
to stop down the telescope to make sure the photomultiplier tube operates
well below saturation. A counter with eight seven segment LED displays
is ideal because it allows displaying up to 99,999,999 counts per
integration period and is easily visible in the dark. Also a 10 MHz
counter is capable of triggering on 10 MHz or 100 nanosecond wide
pulses so again dead time will be minimized. If a computer interface
is used, the counter part should have the same minimum specifications
as the 10 MHz counter. Figure 28 shows a block diagram of a typical
photon counting system.

Figure 28
Photon Counting Block Diagram
Pulse
Conditioner
LeCroy MVL 100
Originally LeCROY made the MVL100 (Amplifier/Comparator) monolithic
integrated circuit that did most of the functions of a pulse conditioner
(see Figure 29).

Figure 29
LeCroy MVL 100 Pulse Conditioner
New
Pulse Conditioner
The MVL100 worked well and only cost about $25. LeCROY no longer makes
these devices and replacements are hybrids costing in the $300 to
$500 range. That is the reason for the development of the pulse conditioner
shown in Figure 30.

Figure 30
New Pulse Conditioner Schematic
Circuit
Description
Video
Amplifier
Integrated circuit U1 (14 pin DIP) is a high-speed video amplifier
(LM733CN) made by National Semiconductor. It has a bandwidth of 120
MHz with selectable gains of 10, 100, and 400. The device operates
from + 5 VDC and - 5 VDC with a total current requirement of less
than 30 mA.
As
used in this circuit, the device is programmed for a gain of 100 (specifications
indicate the gain will be between 80 and 120). This is accomplished
by connecting pin 3 to pin 12. The output pulse from the photomultiplier
tube is terminated in R1 (51 ohm). R2 provides protection from reflections
of the high speed signal. The "IN 2" (pin 1) input is used as the
signal input and causes the output to be inverted (positive going).
The "IN 1" (pin 14) is connected to ground.
Power
is applied to pins 10 (+5 VDC) and 5 (-5 VDC). The 0.1 uF bypass capacitors
are very important and must be used. They must also be close to power
pins 5 and 14. The output of U1 is differential. The "OUT 2" (pin
7) is terminated, through a 0.1 uF capacitor, with a 1 k ohm resistor.
The "OUT 1" provides an amplified positive going pulse to the low-pass
filter. The purpose of the low-pass filter is to reduce high frequency
harmonics in the pulse.
Voltage
Comparator
Integrated circuit U2 (14 pin DIP) is a high-speed voltage comparator
(LM710CN) made by National Semiconductor. Because not all pulses coming
from the photomultiplier tube are due to photons (some are produced
by thermal emission) it is desirable to select only the ones due to
photons. The pulse produced by a photon goes through all the stages
of amplification and attains the highest gain or amplitude.
Some
thermal electrons also may go through all stages but many are emitted
from lower potential stages and produce pulses of lesser amplitude.
By using a comparator with an adjustable threshold voltage, the lower-level
pulses can be eliminated. The other thermal electrons going through
all stages produce the measured dark counts. U2 operates from +12
VDC (pin 11) and -5 VDC (pin 6) supplies.
Total
current for each voltage is less than 10 mA. Again, the 0.1 uF bypass
capacitors are very important. The pulse from the low-pass filter
is routed to the "+" input (pin 3) of U2. Because of the high input
impedance of U2 it is necessary to have resistor R4 (1 k ohms) terminate
the signal. The "-" input to U2 (pin 4) is used for the comparison
voltage input. Any input pulses to pin 3 that equal or exceed the
voltage on pin 4 will cause the output of U2 to go low. The threshold
voltage is produced from the + 5 VDC supply and voltage divider R7
(100k ohms) and multi-turn potentiometer R6 (1k ohm). Typical threshold
voltage on pin 4 of U2 is 20 mV. This is a good starting point for
determining a more precise threshold voltage. The output of U2 is
on pin 9.
Monostable
Multivibrator
Despite its awesome name, the monostable multivibrator is a very simple
and easy-to-use device. Integrated circuit U3 (14 pin DIP) is a monostable
multivibrator, or one-shot, (74121) made by various manufactures.
The purpose of U3 is to form a constant pulse width for each input
pulse and produce two TTL level outputs. The output pulse width is
set by resistor R8 (2k ohms) and capacitor C2 (150 pF) and is about
150 nanoseconds. The device is not re-triggerable, meaning if a second
trigger is applied before the output pulse is finished, it will be
ignored.
With
the input to U3 connected to pin 4 (A2) and pins 3 (A1) and 5 (B)
tied to Vcc, the device will produce an output pulse when the input
is a negative going transition. Only +5 VDC (pin 14) is required for
U3 and the supply current is less than 40 mA. The outputs are complementary.
Pin 6 (Q) produces a positive going pulse while pin 1 (Q NOT) produces
a negative going pulse. Either of the pulses can be used with a counter
because counters count either positive or negative transitions and
both outputs produce one of each per pulse.
Note:
If there is sufficient interest an order of printed wiring boards
can be made for both the pulse conditioner and high voltage power
supply. A basic set of parts could also be included. For more information,
contact Jeff Hopkins at phxjeff@hposoft.com.
Frequency/Pulse
Counter
Introduction
Most any electronic counter will work. It should be able to count
at a minimum of 10 MHz, however. Input should allow TTL pulses to
be counted. The display should have a minimum of 8 LED displays. Surplus
counters can sometimes be purchased for $100 or less.
Check:
http://www.bgmicro.com/
If
you wish, you can build your own counter and save money.
Figure
31 shows a basic counter schematic of an easy-to-build (about $100)
Intersil ICM7226AEV/Kit 10 MHz Universal (Pulse) Counter kit. In addition
to buying the kit, it is suggested that eight right angle sockets
be purchased so as to allow the display to be positioned at right
angles to the printed wiring board. This makes mounting the board
and display much easier. While Intersil no longer makes this, searches
may turn up surplus kits.

Figure 31
10 MHz Pulse Counter