Part
VII
Preliminary System Calibration
Introduction
Before taking the photometry equipment to the telescope some preliminary
calibration must be done. Normally these need be done just once. While
it is suggested to start with a high voltage of -950 VDC, once the signal-to-noise
calibration is set, you can experiment with other settings. If you decide
on a different high voltage setting, you may wish to repeat the signal-to-noise
calibration, however, unless the high voltage has been changed by more
than 10 volts, it may not be necessary.
Signal
-to-Noise Adjustment
There are two variables to consider with a digital photon-counting system.
One is the high voltage on the photomultiplier tube and the other is the
threshold voltage in the pulse conditioning circuit. The optimum settings
for these voltages are when the best Signal-to-Noise or S/N ratio is obtained.
This is a point where the net number of counts per second (e.g., with
and without a fixed amount of light) has the largest difference, i.e.,
the ratio of light counts to dark counts. If you do not wish to do this
calibration, for a 1P21 or equivalent photomultiplier tube, a good threshold
voltage is -9.5 VDC for the MVL100 circuit and 20 millivolts DC for the
new pulse conditioner circuit, with the high voltage set at -950 VDC or
the voltage determined from the next section. Let the tube sit in the
dark with the voltage applied for about 30 minutes to let it stabilize.
The ambient temperature should also be stablized. If the temperature varies
by more than a few degrees during the calibration, it may be best to wait
and do the calibration once the temperature has stablized. A cooler temperature
is preferred to a warmer temperature. Refer to Figure 32 for a test set-up
to do this calibration.

Figure 32
Signal-to-Noise Calibration
Setup
1. With
the Vhv set at -900 VDC and the Vth set at -10.0 VDC (MVL 100) or 10 millivolts
DC (for new circuit), adjust the light intensity for about 5000 counts
per second.
2. Set
Vth to -6.0 VDC (for MVL100) or 5.0 millivolts DC (for new circuit).
3. Set
Vhv to -600 VDC.
4. Record
the counts per second with (CL) and without (CD) the light source.
5. Increment
the Vhv in 50 volt increments, repeat step 4 after each increment, until
-1050 VDC is reached.
6. Next,
increment Vth in 0.5 VDC or 5 millivolts DC increments while repeating
steps 3, 4 and 5 at each increment.
7. Make
a graph of the results. The vertical axis should be the Log10 of the counts.
The horizontal axis should be the Vth settings.
8. The
resulting plots should be a set of curves (CL and CD) for each High Voltage
setting.
9. The
best S/N is with Vhv and Vth settings that produce the largest difference
between the CL and CD curves.
10.
Note these voltages and adjust the system for them. This will give the
best performance.
What
High Voltage Setting to Use
The higher the high voltage, up to about -1,000 VDC, applied to a photomultiplier
tube, the more sensitive the tube is. Higher voltages will also create
more dark current (noise or dark counts). A lot depends on the tube being
used. If more sensitivity is required, then a higher voltage may be considered.
For most photon counting use, -950 VDC is a good value. If the signal-to-noise
calibration has been done, use the high voltage determined by that.
Note:
The system zero points and dead time will change with a high voltage change.
The
following is optional and discussed for those wishing to understand their
photometry system better.
If you
wish to determine what the optimum high voltage is you for your photomultiplier
tube you will need to perform the following experiment. You will need
a light-tight box with a small wheat type bulb or LED to act as a constant
light source. See Figure 29 for a suggested arrangement. The photomultiplier
tube with socket and voltage divider you plan to use in the final equipment
along with a magnetic shield should be used. With no light on in the box,
set the tube's high voltage for -1,000 VDC and let the tube "cook" for
about 30 minutes. This will stabilize the noise counts.
Note:
It is important that the ambient temperature be constant as it will affect
the sensitivity and noise counts. Slight temperature variations (a few
degrees) will not be much of a problem, but large swings (10's of degrees)
may have significant impact and should be avoided.
Next,
adjust the high voltage to your starting voltage, usually the lowest you
want to check. Now set the brightness of the lamp or LED to get counts
in the tens of thousands. The counts will vary slightly due to system
noise, but should average close to each other for a constant high voltage
and a 10 second integration. Make a table of high voltage vs. counts.
You can also prepare a table that varies the high voltage with no light.
This will show the system noise vs. high voltage. From the tables compare
sensitivity vs. noise and decide what high voltage you think is best.
Once
the desired voltage has been determined the power supply should be regulated
±1.0 VDC or better. A digital readout of the high voltage is helpful for
keeping an eye on it. Figure 33 shows a typical plot of high voltage (-900
VDC to -1,000 VDC) vs. sensitivity. As can be seen, a change of the 100
VDC can produce nearly a 0.3 magnitude change.

Figure 33
High Voltage vs. Photomultiplier Tube Sensitivity